EasyManuals Logo

Keysight FieldFox A Series User Manual

Keysight FieldFox A Series
626 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #524 background imageLoading...
Page #524 background image
524 Keysight N9927-90001 User’s Guide
System Settings
CAT/NA
CAT/NA
Choose the CAT/NA settings for SnP file save, default factory error corrections,
extended cal default. See also “> What Data Processing functions are
included in saved SNP Data:” on page 551.
How to select CAT/NA defaults:
—Press System 7
—Then Preferences
—Then Preferences
Then use arrow softkeys or the knob to select CAT/NA
—Then Edit
for each of the following:
—SnP File Save
——Default
——Advanced
Default Factory Error Corr. (N995xA/6xA Only)
——Enhanced Response
——Full 2 Port
Extended Cal Default (N995xA/6xA Only)
——On
——Off (default on models >6 GHz)
—Then Done Edit
(after each change in setting)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight FieldFox A Series and is the answer not in the manual?

Keysight FieldFox A Series Specifications

General IconGeneral
BrandKeysight
ModelFieldFox A Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals