Keysight N9927-90001 User’s Guide 537
System Settings
System Configuration
How to set Power ON
—Press System 7
—Then System Configuration
—Then More
—Then Power ON
Then choose from the following:
Auto
The FieldFox will power ON when a charged battery or the DC
Adapter is inserted.
Manual
The FieldFox will power ON only when the power button is
pressed.
FieldFox Package Installer (N995xA, N996xA, and applies to:
N991xA/2xA/3xA units with serial number prefixes 5607)
The FieldFox Package Installer enables you to download new versions of
firmware from either a USB memory or SD Card. Additionally, the Utility softkey
enables you to backup user files or restore user files.
How to run the Installer
—Press System 7.
—Then System Configuration
.
—Then More
—Then Firmware Update
—Then Start Installer
The Installer software opens after a couple of seconds.
Then the arrows or the knob to scroll to the firmware you want to
install.
—Press Start Install
—“Current Step Progress:” bar and a “Total Progress:” bar are
displayed along with this message: “Copying package. Do not
remove install media.
Before running the Installer to update your firmware, you might want to
back up your user data. Refer to “How to restore files” on page 538.
If no files are displayed, verify that a USB memory or SD Card is installed
and press Refresh List
.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight FieldFox A Series and is the answer not in the manual?

Keysight FieldFox A Series Specifications

General IconGeneral
BrandKeysight
ModelFieldFox A Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals