Repair Guide 6
Keysight N1913/1914A Service Guide 125
Step 4
Overlay the EMI screen onto the key flex circuit, ensuring that
all of the metal tabs are engaged to hold it securely.
Step 5
Fit the display interface board onto the plastic mounting lugs
on the display support molding.
Step 6
Attach the inverter board to the display interface board using
the two screws removed earlier.
Step 7
Connect the inverter board to the display interface board using
the backlight cable assembly.
Note:
The cable must be tucked under the plastic clip to prevent any
fouling.