EasyManua.ls Logo

Keysight N9000B - Scale Range; Ref Position

Keysight N9000B
2530 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3 Short-Range Comms & IoT Mode
3.8 SEM Measurement
Couplings When "Auto Scaling" on page 1040 is ON, this value is automatically determined by the measurement
result. If you set a value manually, "Auto Scaling" on page 1040 automatically changes to OFF
Preset 10.00 dB / Div
State Saved Saved in instrument state
Min 0.10 dB
Max 20 dB
Annotation <value> dB/ left upper of graph
Backwards
Compatibility
SCPI
:DISPlay:SEMask:VIEW[1]:WINDow[1]:TRACe:Y[:SCALe]:PDIVision
Scale Range
Sets the Y-Axis scale range.
Remote Command Replace <meas> with the identifier for the current measurement
:DISPlay:<meas>:WINDow[1]:TRACe:Y[:SCALe]:RANGe <rel_ampl>
:DISPlay:<meas>:WINDow[1]:TRACe:Y[:SCALe]:RANGe?
Example
:DISP:CHP:WIND:TRAC:Y:RANG 100
:DISP:CHP:WIND:TRAC:Y:RANG?
Couplings Coupled to Scale/Div as follows
Scale Range = Scale/Div * 10 (number of divisions)
When you change this value, Auto Scaling automatically changes to OFF
Preset 100 dB
State Saved Saved in instrument state
Min 1
Max 200
Ref Position
Positions the reference level at the top, center, or bottom of the Y Scale display.
Changing the reference position does not change "Ref Value" on page 1038.
Remote
Command
:DISPlay:SEMask:WINDow[1]:TRACe:Y[:SCALe]:RPOSition TOP | CENTer | BOTTom
:DISPlay:SEMask:WINDow[1]:TRACe:Y[:SCALe]:RPOSition?
Example
:DISP:SEM:WIND:TRAC:Y:RPOS CENT
:DISP:SEM:WIND:TRAC:Y:RPOS?
Preset
TOP
Short Range Comms & IoT Mode User's &Programmer's Reference 1039

Table of Contents

Other manuals for Keysight N9000B

Related product manuals