3 Short-Range Comms & IoT Mode
3.2 Modulation Analysis Measurement
Abort (Remote Command Only)
This command is used to stop the current measurement. It aborts the current
measurement as quickly as possible, resets the sweep and trigger systems, and puts
the measurement into an "idle" state. If the analyzer is in the process of aligning
when ABORt is sent, the alignment finishes before the abort function is performed.
So, ABORt does not abort an alignment.
If the analyzer is set for Continuous measurement, it sets up the measurement and
initiates a new data measurement sequence with a new data acquisition (sweep)
taken once the trigger condition is met.
If the analyzer is set for Single measurement, it remains in the "idle" state until an
:INIT:IMM command is received.
Remote Command
:ABORt
Example
:ABOR
Notes If :INITiate:CONTinuous is ON, then a new continuous measurement will start immediately, with sweep
(data acquisition) occurring once the trigger condition has been met
If :INITiate:CONTinuous is OFF, then :INITiate:IMMediate is used to start a single measurement, with
sweep (data acquisition) occurring once the trigger condition has been met
Dependencies For continuous measurement, ABORt is equivalent to the Restart key
Not all measurements support the abort command
Status Bits/OPC
dependencies
The STATus:OPERation register bits 0 through 8 are cleared
The STATus:QUEStionable register bit 9 (INTegrity sum) is cleared
Since all the bits that feed into OPC are cleared by the ABORt, the ABORt will cause the *OPC query to
return true
3.2.9.2 X Scale
This tab accesses controls that enable you to set the horizontal scale parameters.
X Width
Set the width of the X axis, which is displayed for the selected trace.
Remote Command
:DISPlay:EVM:WINDow[1]|2|...|6:X[:SCALe]:WIDTh <real>
:DISPlay:EVM:WINDow[1]|2|...|6:X[:SCALe]:WIDTh?
Example
:DISP:EVM:WIND3:X:WIDT 10e6
:DISP:EVM:WIND3:X:WIDT?
query the X width of the third window
302 Short Range Comms & IoT Mode User's &Programmer's Reference