3 Short-Range Comms & IoT Mode
3.3 LoRa (CSS) Demodulation Measurement
Notes WINDow4: RF envelope window
WINDow5: raw demod waveform window
Couplings If the value is not a multiple of the sample interval, it’ll be adjusted to the closest multiple
For window #4 and #5, the X scale parameters of both windows are coupled when the X axis coupled is
on, and decoupled otherwise
Preset 200 us
State Saved Saved in instrument state
Min 1.000 ns
Max 10.000 s
Ref Position
Sets the reference position of the abscissa of the RF envelope window, the raw
demod waveform window and the frequency drift window, respectively.
Remote
Command
:DISPlay:LORA:WINDow4|5|6:TRACe:X[:SCALe]:RPOSition LEFT | CENTer | RIGHt
:DISPlay:LORA:WINDow4|5|6:TRACe:X[:SCALe]:RPOSition?
Example
:DISP:LORA:WIND4:TRAC:X:RPOS LEFT
:DISP:LORA:WIND4:TRAC:X:RPOS?
Notes WINDow4: RF envelope window
WINDow5: raw demod waveform window
WINDow6: frequency drift window
Preset LEFT
State Saved Saved in instrument state
Range Left|Ctr|Right
X Axis Coupled
Applies only to the RF envelope window and the raw demod waveform window. The
scale/div of both windows will be coupled if this parameter is on and decoupled
otherwise.
Remote
Command
:DISPlay:LORA:WINDow4|5:TRACe:X[:SCALe]:COUPle ON | OFF | 1 | 0
:DISPlay:LORA:WINDow4|5:TRACe:X[:SCALe]:COUPle?
Example
:DISP:LORA:WIND4:TRAC:X:COUP OFF
Sets X Axis coupled to off
Notes WINDow4: RF envelope window
WINDow5: raw demod waveform window
426 Short Range Comms & IoT Mode User's &Programmer's Reference