EasyManua.ls Logo

Keysight N9000B - 8.1.16 Trigger Center Frequency; 8.1.17 Trigger BW

Keysight N9000B
2530 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
8 Trigger
8.1 Trigger
8.1.16 Trigger Center Frequency
Sets the center frequency to be used by the auxiliary receiver for the Auxiliary
Channel I/Q Magnitude trigger.
Remote Command
:TRIGger[:SEQuence]:AIQMag:CENTer <freq>
:TRIGger[:SEQuence]:AIQMag:CENTer?
Example
:TRIG:AIQM:CENT 10 MHz
Notes Trigger CF + 1/2 Trigger BW < Max
Trigger CF - 1/2 Trigger BW > Min
Dependencies Only appears when Aux Channel I/Q Mag is selected as the Trigger Source
Preset 0 Hz
State Saved Saved in instrument state
Range -40 MHz to 40 MHz
Min -40 MHz
Max 40 MHz
8.1.17 Trigger BW
Sets the information bandwidth used by the auxiliary receiver for the Auxiliary
Channel I/Q Magnitude trigger.
Remote Command
:TRIGger[:SEQuence]:AIQMag:BANDwidth <freq>
:TRIGger[:SEQuence]:AIQMag:BANDwidth?
Example
:TRIG:AIQM:BAND 8 MHz
Notes The combined sample rate for the main and auxiliary receivers cannot exceed 100 MSa/sec. The
bandwidth available to Trigger BW is limited to what is available after the main receiver's bandwidth
(Info BW, sometimes pre-FFT BW) is set. Because of this limitation, the Max is not always achievable
The combination of "Trigger Center Frequency" on page 2357 and Trigger BW is also limited:
Trigger CF + 1/2 Trigger BW < Max
Trigger CF - 1/2 Trigger BW > Min
Dependencies Only appears when Aux Channel I/Q Mag is selected as the Trigger Source
Preset Bandwidth option dependent:
No Opt: 10 MHz
Opt B25: 25 MHz
Opt S40: 40 MHz
2357 Short Range Comms & IoT Mode User's &Programmer's Reference

Table of Contents

Other manuals for Keysight N9000B

Related product manuals