3 Short-Range Comms & IoT Mode
3.4 HRP UWB Demodulation Measurement
:HUWB:RSF:ZGAP:NUMB?
Preset
64
State Saved Saved in instrument state
Min
0
Max
64
MMRS Symbol Repetitions
Sets the number of MMRS symbol repetitions within one RSF.
Remote
Command
[:SENSe]:HUWB:RSF:MMRS:SREPetition S32 | S40 | S48 | S64 | S128 | S256
[:SENSe]:HUWB:RSF:MMRS:SREPetition?
Example
:HUWB:RSF:MMRS:SREP S64
Preset
S32
State Saved Saved in instrument state
Range
32|40|48|64|128|256
Number of RSFs
Sets the number of ranging sequence fragments (RSFs). Value 0 means no RSFs
exist.
Remote Command
[:SENSe]:HUWB:RSF:NUMBer N0 | N1 | N2 | N4 | N8
[:SENSe]:HUWB:RSF:NUMBer?
Example
:HUWB:RSF:NUMB N8
Preset
N0
State Saved Saved in instrument state
Range
0|1|2|4|8
STS/RIF Parameters
Number of RIFs (RIF only)
Sets the number of ranging integrity fragments (RIFs). Value 0 means no RIFs exist.
Remote Command
[:SENSe]:HUWB:RIF:NUMBer N0 | N1 | N2 | N4 | N8
[:SENSe]:HUWB:RIF:NUMBer?
Example
:HUWB:RIF:NUMB N8