3 Short-Range Comms & IoT Mode
3.3 LoRa (CSS) Demodulation Measurement
AF Res BW Auto
Remote
Command
[:SENSe]:LORA:AFSPectrum:BANDwidth:AUTO ON | OFF | 1 | 0
[:SENSe]:LORA:AFSPectrum:BANDwidth:AUTO?
Example
:LORA:AFSP:BAND:AUTO ON
:LORA:AFSP:BAND:AUTO?
Preset
AUTO
3.3.5 Display
The Display key opens the Display Menu, which enables you to configure display
items for the current Mode, Measurement View or Window.
3.3.5.1 Meas Display
The Meas Display tab contains controls for setting up the display for the current
Measurement, View or Window.
Metric Settings
This control accesses a menu that includes a 1-of-N menu that enables you to
control which modulation magnitude metrics are displayed. If the modulation
magnitude is set to Peak+ Only, Pk-Pk/2 Only, or RMS Only, the other modulation
magnitude metrics are displayed as “---“ in the metrics window.
Remote
Command
:DISPlay:LORA:VIEW:METRics:MMAGnitude ALL | PPK | PNPK | RMS | RMSRatio
:DISPlay:LORA:VIEW:METRics:MMAGnitude?
Example
:DISP:LORA:VIEW:METR:MMAG PPK
Preset
ALL
State Saved Saved in instrument state
Normal
The Peak +, Peak -, Pk-Pk/2 and RMS results are displayed on the metrics window.
Example
:DISP:LORA:VIEW:METR:MMAG ALL
370 Short Range Comms & IoT Mode User's &Programmer's Reference