2 User Interface
2.15 Status Bar
If the display fills up, scrolling is enabled just as in other X-Series Multi-touch UI
displays.
The Status dialog automatically refreshes as new messages and conditions occur.
At the bottom of the screen is a Clear Message Queue button. This button clears all
errors in all error queues.
Note the following:
Clear Message Queue does not affect the current status conditions
Mode Preset does not clear the message queue
Restore System Defaults (Super Preset) will clear all message queues
*CLS only clears the queue if it is sent remotely and *RST does not affect any
error queue
Switching Modes does not affect any error queues
See "More Information" on page 146
Remote
Command
:SYSTem:ERRor[:NEXT]?
Example
:SYST:ERR?
Notes The return string has the format:
Short Range Comms & IoT Mode User's &Programmer's Reference 145