3 Spectrum Analyzer Mode
3.9 SEM Measurement
Measurement µW Path Control Auto behavior
Modulation
Analysis
Use Standard Path unless tuned frequency > 3.6GHz and IFBW > 15MHz, in
which case choose Full Bypass if conditions warrant(FBP Option is available
and “Allow Full Bypass in Auto” is On ), otherwise choose Preselector Bypass
Monitor Spectrum Always Standard Path
IQ Waveform Use Standard Path unless tuned frequency > 3.6GHz and IFBW > 15MHz, in
which case choose Full Bypass if conditions warrant(FBP Option is available
and “Allow Full Bypass in Auto” is On ), otherwise choose Preselector Bypass
Channel Power Always Standard Path
Occupied BW Always Standard Path
CCDF Use Standard Path unless tuned frequency > 3.6GHz and IFBW > 15MHz, in
which case choose Preselector Bypass
ACP Always Standard Path
SEM Always Standard Path
Spurious
Emissions
Always Standard Path
Transmit On|Off
Power
Use Standard Path unless tuned frequency > 3.6GHz and Info BW > 15MHz, in
which case choose Preselector Bypass
Channel Quality Mode
Measurement µW Path Control Auto behavior
Group Delay Use Standard Path unless tuned frequency > 3.6GHz and IFBW > 15MHz, in
which case choose Full Bypass if conditions warrant(FBP Option is available
and “Allow Full Bypass in Auto” is On ), otherwise choose Preselector Bypass
Monitor Spectrum Always Standard Path
IQ Waveform Use Standard Path unless tuned frequency > 3.6GHz and IFBW > 15MHz, in
which case choose Preselector Bypass
CCDF Use Standard Path unless tuned frequency > 3.6GHz and IFBW > 15MHz, in
which case choose Preselector Bypass
Remote Command
[:SENSe]:POWer[:RF]:MW:PATH:AUTO ON | OFF | 1 | 0
[:SENSe]:POWer[:RF]:MW:PATH:AUTO?
Example
:POW:MW:PATH:AUTO ON
:POW:MW:PATH:AUTO?
Dependencies Only appears in VMA, WLAN, 5G NR and CQM Modes
Couplings See "µW Path Control Auto" on page 1591 above
Preset
ON
Range
ON|OFF