Keysight N9010B - 8.1.18 X Axis Relative to Trigger

Keysight N9010B
2852 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
8 Trigger
8.1 Trigger
Remote Command
:TRIGger[:SEQuence]:AIQMag:BANDwidth <freq>
:TRIGger[:SEQuence]:AIQMag:BANDwidth?
Example
:TRIG:AIQM:BAND 8 MHz
Notes The combined sample rate for the main and auxiliary receivers cannot exceed 100 MSa/sec. The
bandwidth available to Trigger BW is limited to what is available after the main receiver's bandwidth
(Info BW, sometimes pre-FFT BW) is set. Because of this limitation, the Max is not always achievable
The combination of "Trigger Center Frequency" on page 2676 and Trigger BW is also limited:
Trigger CF + 1/2 Trigger BW < Max
Trigger CF - 1/2 Trigger BW > Min
Dependencies Only appears when Aux Channel I/Q Mag is selected as the Trigger Source
Preset Bandwidth option dependent:
No Opt: 10 MHz
Opt B25: 25 MHz
Opt S40: 40 MHz
State Saved Saved in instrument state
Range 10 Hz to Maximum
Min 10 Hz
Max Bandwidth option & I/Q input path-dependent:
No Opt, I or Q Only: 10 MHz, I+jQ: 20 MHz
Opt B25, I or Q Only: 25 MHz, I+jQ: 50 MHz
Opt S40, I or Q Only: 40 MHz, I+jQ: 80 MHz
8.1.18 X Axis Relative to Trigger
For triggers that support Trigger Delay, it is beneficial that in time-domain displays
(like Zero Span), X-axis values should be referenced to the Trigger point, that is, the
zero value on the X Axis should be wherever the trigger point appears on the X Axis
(including to the left or right of the X Axis). For negative trigger delays this means
the zero point can actually appear on the X Axis.
Traditionally the zero point is at the left edge of the X Axis, and this behavior is
retained for backwards compatibility, but if you turn on X Axis Relative to Trigger,
the zero point moves to the trigger point and all X Axis values change to be relative
to the trigger point.
When the X Axis Relative to Trigger switch is On, the trigger point on the X-axis will
be marked with a vertical line and an annotation of “TRIG”. Additionally, when the
2677 Spectrum Analyzer Mode User's &Programmer's Reference

Table of Contents