3 Spectrum Analyzer Mode
3.13 Phase and Amplitude vs Time Measurement
Meas Transient
Sets the time between the end of the "Meas Interval" on page 2014 and the end of
the segment.
Remote
Command
[:SENSe]:PAVTime:SEGMents:TRANsient <time>
[:SENSe]:PAVTime:SEGMents:TRANsient?
Example
:PAVT:SEGM:TRAN 1 us
:PAVT:SEGM:TRAN?
Couplings Meas Transient must be >= 0 s and less than (Max Capture Length/Segments) - Meas Offset - Meas
Interval
Preset 0
State Saved Saved in instrument state
Min/Max Min = 0
Max = (Max Capture Length/Segments) - Meas Offset - Meas Interval
Measure Frequency Error
Measures the frequency error of the input signal at the center frequency.
This is an "immediate action" function, that is, it executes once, when the control is
pressed; the trigger is effectively Free Run, because the acquisition occurs
immediately regardless of the instrument trigger setting.
Remote Command
[:SENSe]:PAVTime:FERRor:IMMediate
Example
:PAVT:FERR:IMM
Status Bits/OPC depend-
encies
Non-overlapped
Frequency Error
The offset of the input signal from the Center Frequency. The frequency error can be
determined by the instrument using "Measure Frequency Error" on page 2015, or is
can be supplied by the user.
Remote Command
[:SENSe]:PAVTime:FERRor <freq>
[:SENSe]:PAVTime:FERRor?
Example
:PAVT:FERR 10 MHz
:PAVT:FERR?
Preset 0 Hz
Spectrum Analyzer Mode User's &Programmer's Reference 2015

Table of Contents

Other manuals for Keysight N9010B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight N9010B and is the answer not in the manual?

Keysight N9010B Specifications

General IconGeneral
BrandKeysight
ModelN9010B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals