3 Spectrum Analyzer Mode
3.3 Channel Power Measurement
Measure Trace
Specifies which trace’s scalar results are displayed in the Metrics window, and
retrieved by sending a :READ or :FETCh query:
Trace 1
Trace 2
Trace 3
Remote Command
:CALCulate:<meas>:MTRace TRACe1 | TRACe2 | TRACe3
:CALCulate:<meas>:MTRace?
<meas> is the identifier for the current measurement; any one of CHPower | ACPower |
OBWidth | SEMask | SPURious | PVTime
Example Channel Power
:CALC:CHP:MTR TRAC1
:CALC:CHP:MTR?
Dependencies In the ACP measurement, this control is grayed-out when Meas Method is set to RBW or FAST, and
only Trace 1 is enabled
Preset
TRACe1
State Saved No
Range Trace 1 | Trace 2 | Trace 3