4 System
4.6 Alignments
Query Temperature of Last Align Now 30 MHz 3.6 MHz
Remote
Command
:CALibration:TEMPerature:RFPSelector:LRADiated?
Example
:CAL:TEMP:RFPS:LRAD?
Notes Value is in degrees Centigrade at which the last successful Align Now, 30MHz 3.6GHz was executed
State Saved No
Query Next Scheduled Alignment Time
Remote
Command
:CALibration:RFPSelector:SCHeduler:TIME:NEXT?
Returns data using the following format:
YYYY/MM/DD; HH:MM:SS
Example
:CAL:RFPS:SCH:TIME:NEXT?
Notes The next run time will be updated based on the start date/time and recurrence set by the user
“date” is representation of the date the task will run in the form:
YYYY/MM/DD
where:
YYYY is the four-digit representation of year. (for example, 2009)
MM is the two-digit representation of month. (for example, 01 to 12)
DD is the two-digit representation of the day. (for example, 01 to 28, 29, 30 or 31 depending on the
month and year)
“time” is a representation of the time of day the task will run in the form:
HH:MM:SS
where:
HH is the two-digit representation of the hour in 24-hour format
MM is the two-digit representation of minute
SS is the two-digit representation of seconds
State Saved No
Query Time since Last Align Now External Mixer Path
Remote Command
:CALibration:TIME:INTernal:EMPath?
Example
:CAL:TIME:INT:EMP?
Notes Value in hours since the last successful Align Now External Mixer Path
Dependencies Only appears option MXP is installed
State Saved No