Command Reference
Power Measurement Subsystem
112
Channel Power Auto Range
[:SENSe]:CHPower:METer:RANGe:AUTO OFF|ON|0|1
[:SENSe]:CHPower:METer:RANGe:AUTO?
Toggles the channel power auto range between on and off.
It’s only available when the display mode is set to Meter.
*RST: ON
Example: :CHPower:METer:RANGe:AUTO 0
Key access: Meas > Disp Range > Auto Range
Channel Power Range Top
[:SENSe]:CHPower:METer:RANGe:TOP <value>
[:SENSe]:CHPower:METer:RANGe:TOP?
Set the top amplitude range of the channel power
measurement. it’s only available when the channel power
auto range is set to off.
*RST: –10
Range: Bottom to –10
Example: :CHPower:METer:RANGe:TOP -30
Key access: Meas >Disp Range > Top
Channel Power Range Bottom
[:SENSe]:CHPower:METer:RANGe:BOTTom <value>
[:SENSe]:CHPower:METer:RANGe:TOP?
Set the bottom amplitude range of the channel power
measurement. it’s only available when the channel power
auto range is set to off.
*RST: –70
Range: –140 to Top
Example: :CHPower:METer:RANGe:BOTTom -60
Key access: Meas >Disp Range > Bottom