127
Command Reference
Spectrum Monitor Option Subsystem
Example: :SPECtrogram:UINTerval:STATe ON
Key access: Meas > Update Int.
Spectrogram Palette
[:SENSe]:SPECtrogram:PALette FCOLor|GRAY
[:SENSe]:SPECtrogram:PALette?
This command switches the display palette between Color
and Gray.
Example: :SPECtrogram:PALette GRAY
Key access: Meas > Palette > Color /Gray
Spectrogram Display Mode
[:SENSe]:SPECtrogram:DMODe SPCTrg|TRACe|BOTH
[:SENSe]:SPECtrogram:DMODe?
Use this command to switch the spectrogram display mode
between Spectrogram, Trace and both display.
Range: SPEC, TRACe, BOTH
Example: :SPECtrogram:DMODe BOTH
Key access: Meas > Disp Mode
Spectrogram Marker State
[:SENSe]:SPECtrogram:MARKer[1]|2:STATe
OFF|ON|0|1
[:SENSe]:SPECtrogram:MARKer[1]|2:STATe?
Use this command to turn on/off the marker in current test.
Example: :SPECtrogram:MARKer2:STATe ON
Key access: Marker > State
Spectrogram Display Frequency
[:SENSe]:SPECtrogram:MARKer[1]|2:FREQuency
<value>
[:SENSe]:SPECtrogram:MARKer[1]|2:FREQuency?

Table of Contents