EasyManuals Logo

Keysight N9322C User Manual

Keysight N9322C
223 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #147 background imageLoading...
Page #147 background image
139
Command Reference
Reflection Measurement Option Subsystem
Reflection Measurement Amplitude Scale Per Division
[:SENSe]:CAT:SCALe:DIV <value>
[:SENSe]:CAT:SCALe:DIV?
This command defines the amplitude display scale per
division.
Range: 0.1 to 10
Example: :CAT:SCALe:DIV 5
Key access: Meas > Start Distance
Reflection Measurement Save Trace Data to Memory
[:SENSe]:CAT:TRACe:SAVE
Use this command to save the current trace data to memory.
Example: :CAT:TRACe:SAVE
Key access: Trace > Data >Mem
Reflection Measurement Trace Display Type
[:SENSe]:CAT:TRACe:DISPlay DATA|MEM|DAM
[:SENSe]:CAT:TRACe:DISPlay?
This command toggles the trace display type. Reflection
measurement supports three display type, displaying current
measurement result, displaying the trace data saved in
memory, and display both the data saved in memory and
current sweep result.
Range: DATA, MEM, DAM
Example: :CAT:TRACe:DISPlay MEM
Key access: Trace> Data / Memory / Data&Memory