145
Command Reference
Reflection Measurement Option Subsystem
Get Limits Result in Reflection Measurement
:CALCulate:LLINe[1]|2:FAIL?
This command query the result of limit line or limit in
reflection measurement.
Example: :CALCulate:LLINe1:FAIL?
Define Limit Value in Reflection Measurement
:CALCulate:LLINe[1]|2:DATA
<x-axis>,<ampl>,<connected>{,<x-axis>,<ampl>,<co
nnected>}
:CALCulate:LLINe[1]|2:DATA?
Use this command to define the limit value in reflection
measurement function.
Example: :CALC:LLIN1:DATA
1000000000,-20,0,200000000,-30,1
Key access: Shift > Limit > Limit1 > Edit Limits
Delete Limit Data in Reflection Measurement
:CALCulate:LLINe[1]|2:DELete
This command delete the current defined limit value.
Example: :CALCulate:LLINe1:DELete
Get Limits Result in Reflection Measurement
:CALCulate:LLINe[1]|2:FAIL?
This command query the result of limit line or limit in
reflection measurement mode.
Example: :CALCulate:LLINe1:FAIL?

Table of Contents