EasyManua.ls Logo

Keysight N9322C - Page 172

Keysight N9322C
223 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command Reference
Demodulation Option Subsystem
164
Carrier Frequency Step
[:SENSe]:FREQuency:CARRier:STEP <freq>
[:SENSe]:FREQuency:CARRier:STEP?
This command sets the carrier frequency step of the
demodulation analysis.
Example: :FREQuency:CARRier:STEP 50MHz
Key access: Freq > CarrFreq Step
Auto Carrier Frequency State
[:SENSe]:FREQuency:CARRier:AUTO OFF|ON|1|0
[:SENSe]:FREQuency:CARRier:AUTO?
Use this command to turn on/off auto carrier frequency
detection.
Example: :FREQuency:CARRier:AUTO ON
Key access: Freq > Auto CarrFreq
Attenuation
[:SENSe]:POWer[:RF]:ATTenuation <value>
[:SENSe]:POWer[:RF]:ATTenuation?
This command sets the RF attenuation value.
Range: 0 to 51 dB
Example: :POWer:ATTenuation 20dB
[:SENSe]:POWer[:RF]:ATTenuation:AUTO OFF|ON|0|1
[:SENSe]:POWer[:RF]:ATTenuation:AUTO?
Use this command to turn on/off auto attenuation status.
Example: :POWer:ATTenuation:AUTO ON
Key access: Amptd > Attenuation

Table of Contents

Related product manuals