Command Reference
Demodulation Option Subsystem
164
Carrier Frequency Step
[:SENSe]:FREQuency:CARRier:STEP <freq>
[:SENSe]:FREQuency:CARRier:STEP?
This command sets the carrier frequency step of the
demodulation analysis.
Example: :FREQuency:CARRier:STEP 50MHz
Key access: Freq > CarrFreq Step
Auto Carrier Frequency State
[:SENSe]:FREQuency:CARRier:AUTO OFF|ON|1|0
[:SENSe]:FREQuency:CARRier:AUTO?
Use this command to turn on/off auto carrier frequency
detection.
Example: :FREQuency:CARRier:AUTO ON
Key access: Freq > Auto CarrFreq
Attenuation
[:SENSe]:POWer[:RF]:ATTenuation <value>
[:SENSe]:POWer[:RF]:ATTenuation?
This command sets the RF attenuation value.
Range: 0 to 51 dB
Example: :POWer:ATTenuation 20dB
[:SENSe]:POWer[:RF]:ATTenuation:AUTO OFF|ON|0|1
[:SENSe]:POWer[:RF]:ATTenuation:AUTO?
Use this command to turn on/off auto attenuation status.
Example: :POWer:ATTenuation:AUTO ON
Key access: Amptd > Attenuation