Page #193 background image
185
Command Reference
Demodulation Option Subsystem
ASK Carrier Power Upper Limit
:CALCulate:DMA:LIMit:POWer:UPPer <value>
:CALCulate:DMA:LIMit:POWer:UPPer?
Use this command to specify the carrier power upper limit
value of ASK demodulation analysis.
Range: –100 to 30 dBm
Example: :CALCulate:DMA:LIMit:POWer:UPPer 20dBm
Key access: Shift > Limit > CarrPwr Upper
ASK Depth Upper Limit
:CALCulate:ASK:LIMit:AMDepth:UPPer <real>
:CALCulate:ASK:LIMit:AMDepth:UPPer?
Use this command to specify the depth upper limit value of
ASK demodulation analysis.
Range: 0.1% to 100%
Example: :CALCulate:ASK:LIMit:AMDepth:UPPer 20
Key access: Shift > Limit > ASK Depth Up
ASK Depth Lower Limit
:CALCulate:ASK:LIMit:AMDepth:LOWer <real>
:CALCulate:ASK:LIMit:AMDepth:LOWer?
Use this command to specify the depth lower limit value of
ASK demodulation analysis.
Range: 0.1% to 100%
Example: :CALCulate:ASK:LIMit:AMDepth:LOWer 20
Key access: Shift > Limit > ASK Depth Lower Limit