191
Command Reference
Demodulation Option Subsystem
FSK Filter Symbols
[:SENSe]:DMA:RADio:STANdard:FILTer:SYMBol <
integer >
[:SENSe]:DMA:RADio:STANdard:FILTer:SYMBol?
Use this command to specify FSK filter symbols.
Range: 2 to 25
Example: :DMA:RADio:STANdard:FILTer:SYMBol 4
Key access: Meas > Filter Setup > FilterSymbols
FSK Alpha/BbT
[:SENSe]:DMA:RADio:STANdard:ALPHa <real>
[:SENSe]:DMA:RADio:STANdard:ALPHa?
Use this command to specify the filter Alpha/Bbt value.
Range: 0.2 to 1
Example: :DMA:RADio:STANdard:ALPHa 0.2
Key access: Meas > Filter Setup > Alpha/BbT
FSK Measurement Length
[:SENSe]:DMA:MEASurement:LENGth <integer>
[:SENSe]:DMA:MEASurement:LENGth?
Use this command to specify measurement length of FSK
demodulation analysis.
Range: 20 to 1100
Example: :DMA:MEASurement:LENGth 200
Key access: Meas > Meas Length