Command Reference
Power Meter Option Subsystem
200
Key access: Meas > Pause/ Run
Switch Power Range
[:SENSe]:POWer:AC:RANGe UPPer|LOWer
[:SENSe]:POWer:AC:RANGe?
This command toggles the measure power range between
upper and lower. It’s only available for U2000 series average
power sensor.
Example: :POWer:AC:RANGe UPPer
Key access: Meas > General Setup > Power Range > Range > Upper / Lower
Set Power Auto Range State
[:SENSe]:POWer:AC:RANGe:AUTO OFF|ON|0|1
[:SENSe]:POWer:AC:RANGe:AUTO?
This command turns on/off power auto range state. N9322C
will choose the power range according to current
measurement power.
Example: :POWer:AC:RANGe:AUTO 1
Key access: Meas > General Setup > Power Range > Range > Upper / Lower
Set Power Meter Center Frequency
[:SENSe]:FREQuency[:CW] < freq >
[:SENSe]:FREQuency[:CW]?
This command defines the center frequency of power meter
measurement.
Range: 1 kHz to 90 GHz
Example: :SENSe:FREQuency:CW 100MHz
Key access: Meas > General Setup > Freq