205
Command Reference
Power Meter Option Subsystem
This command defines trace Y scale value of peak power
sensor.
Example: :SENSe:TRACe:Y:TIME 1
Key access: Meas > Peak Setup > Trace Setup > Y Max
Set Gate Start Time
[:SENSe]:SWEep[1]:OFFSet:TIME <value>
[:SENSe]:SWEep[1]|:OFFSet:TIME?
This command specify the gate start time of peak power
meter.
Range: –1 to 1 s
Example: :SWEep1:OFFSet:TIME 1
Key access: Meas > Peak Setup > Gate Setup > Start
Set Gate Length Time
[:SENSe]:SWEep[1]:TIME <value>
[:SENSe]:SWEep[1]:TIME?
This command defines the gate length time of peak power
meter measurement.
Range: 0 to 1 s
Example: :SENSe:SWEep1:TIME 1
Key access: Meas > Peak Setup > Gate Setup > Length
Set Display Resolution
:DISPlay[:WINDow]:NUMeric:RESolution T1|T2|T3|T4
:DISPlay[:WINDow]:NUMeric:RESolution?
This command toggles the readout display resolution
between T1 to T4. The display resolution T4 can achieve the
best resolution.
Example: :DISPlay:WINDow:NUMeric:RESolution T4
Key access: Meas > Meas Disp > Resolution