57
Command Reference
System Subsystem
Query Time Base Calibration Result
:SYSTem:CALibration:TBASe:FREQuency?
Use this command to query the time base calibration result.
Calibration Amplitude
:CALibration
Use this command to trigger the amplitude calibration.
Connect the amplitude calibration reference output port to
the RF input port, and turn on the amplitude calibration
reference output
Example: :CALibration
Key access: System > More > Service > Calibration > Amp Alignment
Amplitude Calibration Reference Signal Output
:CALibration:SOURce:STATe ON|OFF|1|0:
:CALibration:SOURce:STATe?
Use this command to turn on/off the amplitude calibration
reference output.
Example: :CALibration
Key access: System > More > Service > Calibration > Cal Out On/off
Query Amplitude Calibration Result
:SYSTem:CALibration:AMPLitude:STATe?
Use this command to query the amplitude calibration result
BNC Port Input Type Setting
:SYSTem:PORT:EXTInput REF|TRIGger
:SYSTem:PORT:EXTInput?
Toggles the external input between 10 MHz reference signal
and TTL signal.
Key access: System > More > Port Setting > Ext Input