EasyManua.ls Logo

Keysight N9322C - Memory Subsystem

Keysight N9322C
223 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
57
Command Reference
System Subsystem
Query Time Base Calibration Result
:SYSTem:CALibration:TBASe:FREQuency?
Use this command to query the time base calibration result.
Calibration Amplitude
:CALibration
Use this command to trigger the amplitude calibration.
Connect the amplitude calibration reference output port to
the RF input port, and turn on the amplitude calibration
reference output
Example: :CALibration
Key access: System > More > Service > Calibration > Amp Alignment
Amplitude Calibration Reference Signal Output
:CALibration:SOURce:STATe ON|OFF|1|0:
:CALibration:SOURce:STATe?
Use this command to turn on/off the amplitude calibration
reference output.
Example: :CALibration
Key access: System > More > Service > Calibration > Cal Out On/off
Query Amplitude Calibration Result
:SYSTem:CALibration:AMPLitude:STATe?
Use this command to query the amplitude calibration result
BNC Port Input Type Setting
:SYSTem:PORT:EXTInput REF|TRIGger
:SYSTem:PORT:EXTInput?
Toggles the external input between 10 MHz reference signal
and TTL signal.
Key access: System > More > Port Setting > Ext Input

Table of Contents

Related product manuals