91
Command Reference
Calculate Subsystem
Limit Line Y-axis Value
:CALCulate:LLINe[1]:Y <value>
:CALCulate:LLINe[1]:Y?
Sets the Y- axis value of a limit line. Limit line Y- axis value
is set independently and is not affected by the X- axis units.
*RST: 0 dBm
Example: :CALCulate:LLINe:Y -20dBm
Key access: Shift > Limit> Limit 1 > Limit Line
Limit Line X-axis Value
:CALCulate:LLINe:CONTrol:DOMain FREQuency|TIME
:CALCulate:LLINe:CONTrol:DOMain?
Toggles the limit line X- axis value between frequency and
time.
Example: :CALCulate:LLINe:CONTrol:DOMain FREQuency
Key access: Shift > Limit > X Axis Units
Limit State
:CALCulate:LLINe[1]|2:DISPlay OFF|ON|0|1
:CALCulate:LLINe[1]|2:DISPlay?
Toggles the limits state between on and off.
Key access: Shift > Limit > Limit 1> Limit On/Off
Query Current Limits Result
:CALCulate:LLINe[1]|2:FAIL?
This query returns the limits pass/failed result. If the test
result fail, this command will get result 1. If the test result
pass the limit, it will get result 0.
Example: :CALCulate:LLINe:FAIL?