180 Appendix E
Caring for Connectors
Principles of Microwave Connector Care
Gaging
DO DO NOT
Clean and zero the gage before
using.
Use correct gage type.
Use correct end of calibration
block.
Gage all connectors before first
use.
Use an out-of-spec connector.
Making Connections
DO DO NOT
Align connectors carefully.
Make preliminary connection
lightly.
Turn connector nut only to tighten.
Use a torque wrench for final
connection.
Apply bending force to
connection.
Overtighten preliminary
connection.
Twist or screw in connectors.
Tighten past “break” point of
torque wrench.
Table E-4 Principles of Microwave Connector Care