Chapter 2 21
Test Descriptions
10 MHz Out Frequency Reference Adjustment
Step 4. Wait for the frequency counter reading to stabilize and ensure that the 
adjusted value is within the published specification. Repeat the 
adjustment procedure until the correct value is obtained.
Step 5. Press 
Save from the Time Base sub-menu to ensure that the values are 
stored in memory.
NOTE It is recommended that you preset the NFA and repeat the 10 MHz 
Frequency Reference Accuracy Test (see page  12) to verify the adjusted 
DAC settings.