54 Chapter 2
Test Descriptions
Noise Source Supply Accuracy
Step 8. Set the noise source on (press the System key, then More, More and
Service. Press Noise Source On). Press Enter Password then enter the
service password -2010. Press
Enter then Service and Noise Source On).
Step 9. Enter the digital multimeter reading in the Test Record as the Noise
Source Supply On measurement.
Step 10. Press the
Preset key to return to the default Noise Source Supply State.
Step 11. Verify that the measured values are within the published specifications.