1 Performance Verification Overview
20 Keysight U8903B Service Guide
Recommended Test Equipment
The recommended test equipment for performance verification tests is listed below. If the
exact equipment is not available, substitute the calibration standards of equivalent
requirement(s).
Table 1-1 Recommended equipment for performance verification tests
Equipment Recommended model Critical specification
Digital multimeter Keysight 3458A
– DC voltage range of 0 V to 10 V or higher
– AC voltage range of 0 Vrms to 10 Vrms or higher
– 8.5-digit resolution or higher
– DC voltage accuracy of 0.3% or higher
– AC voltage accuracy of 0.02% or higher
Calibrator Fluke 5720A
– DC voltage range of 0 V to 150 V or higher
– AC voltage range of 0 Vrms to 150 Vrms or higher
– Frequency range of 20 Hz to 100 kHz or higher
– DC voltage accuracy of 0.2% or higher
– AC voltage accuracy of 0.02% or higher
Frequency counter
Keysight 53132A Option 010
Keysight 53220A Option 010
– Frequency range of 50 Hz to 80 kHz
– Accuracy of 1 ppm or higher
Oscilloscope
Keysight DSO8064A
Keysight DSO8104A
–
50 Ω terminator – –
Shorting cap (BNC) – –
Computer – GPIB/USB/LAN and APIB connections
BNC cable Keysight U8903B-101 1.2 m
XLR cable Keysight U8903B-102 2 m
BNC-to-banana jack adapter – –
BNC-T adapter – –