EasyManua.ls Logo

Keysight U8903B - Figure 2-4 U8903 B Audio Analyzer Connections for Crosstalk Verification

Keysight U8903B
165 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Performance Verification for Analog Generator 2
Keysight U8903B Service Guide 31
Figure 2-4 U8903B audio analyzer connections for crosstalk verification
2 Configure the U8903B generator output channels as follows:
Connection type: Unbalanced
Output impedance: 600 Ω
Output grounding configuration: Floating ground
Waveform: Sine
Output state: OFF
3 Configure the U8903B analyzer input channels as follows:
Connection type: Unbalanced
Input bandwidth: 90 kHz
Input coupling: DC
Voltage detector type: RMS
Trigger source: Immediate
Multi-channel function: Crosstalk
Sample size: 131072
4 The first test will measure the generator crosstalk from channel 1 to channel 2 in which
the Driving channel is channel 1 and the Measured channel is channel 2.
5 Set the U8903B analyzer Driving channel as the reference channel and the analyzer
input range of Vrms as given in the “Test record for crosstalk verification” on
page 102.
6 Set the U8903B analyzer Measured channel input impedance to 300 Ω and analyzer
input range to 0.32 Vrms.
7 On the U8903B generator, set the Driving channel output amplitude of Vrms as given in
the “Test record for crosstalk verification” on page 102 and the Measured channel
output amplitude to 0 Vrms.
8 Set the U8903B generator Driving and Measured channel frequency of f
DUT
as given in
the “Test record for crosstalk verification” on page 102.
Unbalanced
Output
Channel 2 Channel 1
U8903B audio analyzer
Channel 1
Channel 2
Unbalanced
Input