Performance Verification for Analog Analyzer 3
Keysight U8903B Service Guide 43
Figure 3-3 Function generator to frequency counter and U8903B connection for frequency accuracy
verification
3 Configure the function generator output channel as follows:
Output load: High-Z load
Output polarity: Normal
Waveform function: Sine wave
Amplitude unit = Vrms
Amplitude offset = 0 V
4 Configure the frequency counter input channels as follows:
If frequency counter model is 53220A, set SCPI language mode: 53132A
–Impedance: 1 MΩ
Low pass filter: ON
Input coupling: DC
Level at center of hysteresis window: 0 V
Trigger level percentage of signal peak-to-peak range: 50%
Reference timebase: Internal reference
Measurement function: Channel 1 frequency
Frequency measurement start arm mode: Immediate
Frequency measurement stop arm mode: Timer
5 Configure the U8903B input channels to be tested as follows:
Connection type: Unbalanced
Input bandwidth: 90 kHz
Input coupling: DC
Voltage detector type: RMS
Trigger source: Immediate
Unbalanced input
channel 1, 3, 5, 7
Function generator
Frequency counter
Unbalanced input
channel 2, 4, 6, 8
U8903B audio
analyzer