EasyManua.ls Logo

Keysight U8903B - Figure 3-8 Function Generator to Frequency Counter and U8903 B Connection for Wideband Frequency Accuracy Verification

Keysight U8903B
165 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3 Performance Verification for Analog Analyzer
58 Keysight U8903B Service Guide
2 Connect the function generator outputs to the frequency counter input channel 1 and
2, and to the U8903B unbalanced input channel 1 and 2 as shown in Figure 3-8. Preset
the U8903B, function generator, and frequency counter.
Figure 3-8 Function generator to frequency counter and U8903B connection for wideband frequency
accuracy verification
3 Configure the function generator output channel as follows:
Output load: High-Z load
Output polarity: Normal
Waveform function: Sine wave
Amplitude unit = Vrms
Amplitude offset = 0 V
4 Configure the frequency counter input channels as follows:
If frequency counter model is 53220A, set SCPI language mode: 53132A
–Impedance: 1 MΩ
Low pass filter: ON (for frequency <100 kHz)
Input coupling: DC
Level at center of hysteresis window: 0 V
Trigger level percentage of signal peak-to-peak range: 50%
Reference timebase: Internal reference
Measurement function: Channel 1 frequency
Frequency measurement start arm mode: Immediate
Frequency measurement stop arm mode: Timer
5 Configure the U8903B input channels to be tested as follows:
Connection type: Unbalanced
Input bandwidth: 90 kHz
Unbalanced input
channel 1
Function generator
Frequency counter
Unbalanced input
channel 2
U8903B audio
analyzer