Performance Verification Overview 1
Keysight U8903B Service Guide 21
Test Consideration
For optimum performance, all procedures should comply with the following
recommendations:
Ensure that the calibration ambient temperature is stable and between 18 °C and 28 °C.
Ensure the ambient relative humidity is less than 80%.
Allow at least 30 minutes of warm-up period upon power on.
Use shielded cables only. Keep the cables that connect the test setup as short as
possible.
In the event of a test failure, it is recommended to repeat the test to ensure that
procedural error was not made.
The verification and adjustment procedures are based on the assumption that the
recommended test equipment is being used. Substituting with an alternative test
equipment may require modification of some procedures.