Performance Verification for Digital Generator and Digital Analyzer 4
Keysight U8903B Service Guide 77
– SINE waveform amplitude: 1 FFS
– Digital output state: ON
4 Set the digital output sampling rate to F
SR
as given in “Test record for unbalanced
output level accuracy verification” on page 152.
5 Set the digital output level to V
DUT
as given in “Test record for unbalanced output
level accuracy verification” on page 152.
6 Configure the DSO8064A oscilloscope as follows:
– Timebase scale: 1 / (Sampling rate × 32 × 2)
– Channel 1 vertical scale: Refer to Table 4-8
7 Set the DSO8064A oscilloscope to measure the amplitude of Channel 1 and record the
measurement as V
measure
.
8 Compute the level accuracy using the following equation:
Level Accuracy (dB) = 20
×
Log (V
measure
/ V
DUT
)
9 Repeat step 5 to step 8 for the rest of the V
DUT
as given in “Test record for unbalanced
output level accuracy verification” on page 152 to complete the verification of the
different voltages.
10 Repeat step 4 to step 9 for the rest of the F
SR
as given in “Test record for unbalanced
output level accuracy verification” on page 152 to complete the verification of the
different sampling rates.
Table 4-8 Channel 1 vertical scale settings
V
DUT
(V)
Voltage per division (V)
0.3 0.05
0.5 0.085
0.8 0.15
10.17
1.2 0.2
1.3 0.25
1.5 0.25
1.8 0.3
2.0 0.35
2.3 0.39
2.5 0.42