Performance Verification for Analog Generator 2
Keysight U8903B Service Guide 29
Sine wave frequency accuracy verification
This test verifies the sine wave frequency accuracy of the signal generated by the U8903B
using a frequency counter to ensure that it is within the published specification.
1 Connect the U8903B Unbalanced output channel 1 to the frequency counter channel 1
and Unbalanced output channel 2 to the frequency counter channel 2 as shown in
Figure 2-3. Preset both the U8903B and the frequency counter.
Figure 2-3 U8903B to frequency counter connections for frequency accuracy verification
2 Configure the frequency counter input channels as follows:
If frequency counter model is 53220A, set SCPI language mode: 53132A
–Impedance: 1 MΩ
Low pass filter: ON
Input coupling: DC
Level at center of hysteresis window: 0 V
Trigger level percentage of signal peak-to-peak range: 50%
Reference timebase: Internal reference
Measurement function: Channel 1 frequency
Frequency measurement start arm mode: Immediate
Frequency measurement stop arm mode: Timer
3 Configure the U8903B output channels to be tested as follows:
Connection type: Unbalanced
Output impedance: 600 Ω
Output grounding configuration: Floating ground
Table 2-3 Frequency accuracy specifications
Frequency accuracy specification Equipment used
± (2 ppm + 100 μHz) Frequency counter
Unbalanced output
channel 1
U8903B audio
analyzer
Frequency
counter
Unbalanced output
channel 2