Performance Verification for Digital Generator and Digital Analyzer 4
Keysight U8903B Service Guide 81
Unbalanced output sampling rate accuracy verification
This test measures the sampling rate accuracy on the unbalanced digital audio output of
the U8903B using the Keysight 53132A/53220A frequency counter to ensure that it is
within the published specification.
1 Connect Keysight 53132A/53220A frequency counter input channel 1 to the U8903B
unbalanced digital output channel using the SPDIF or 75 Ω impedance cable.
Figure 4-9 Keysight 53132A/53220A to U8903B unbalanced digital output for unbalanced output
sampling rate accuracy verification
2 Configure the Keysight 53132A/53220A frequency counter as follows:
Preset the equipment
Channel 1 input impedance: 50 Ω
Frequency measurement start arming source: Immediate
Frequency measurement stop arming source: Timer
Frequency arming and measurement gate time: 0.5 s
3 Configure the U8903B digital output channel to be tested as follows:
Preset the U8903B
Digital output type: Unbalanced
Digital output voltage: 2.5 V
Digital output sampling rate: 48 kHz
Digital output AES audio resolution: 24 bits
Digital output function: SINE waveform
SINE waveform frequency: 1 kHz
SINE waveform amplitude: 1 FFS
Digital output state: ON
Table 4-10 Unbalanced output sampling rate accuracy specification
Unbalanced output sampling rate accuracy
specification
Equipment used
±5 ppm Frequency counter
SPDIF or 75 Ω
impedance cable
53132A/53220A
frequency counter
U8903B
unbalanced
digital output