3 Performance Verification for Analog Analyzer
56 Keysight U8903B Service Guide
Figure 3-7 Function generator to U8903B and digital multimeter for wideband sine wave flatness
verification
2 Configure the digital multimeter as follows:
Measurement function: AC voltage
Level Filter: Enable
–NPLC: 10
Number readings: 1 (with auto sample event)
Trigger arming: Hold (suspend)
Trigger mode: Auto
ACV mode: Synchronous sub-sampled
Input Range: Refer to “Test record for wideband sine wave flatness verification
on page 117.
3 Configure the function generator as follows:
Output load: High-Z load
Output polarity: Normal
Waveform function: Sine wave
Amplitude unit: Vrms
Amplitude offset: 0 V
4 Configure the U8903B input channel 1 to be tested as follows:
Connection type: Unbalanced
Input bandwidth: 1.5 MHz
Input coupling: DC
Voltage detector type: RMS
Trigger source: Immediate
Measurement function2: AC Voltage
Sample size: 1048576 for 20 Hz / 524288 for 50 Hz / 262144 for above 50 Hz
Unbalanced input
channel 1 and 2
Function generator
Digital Multimeter
U8903B audio
analyzer