4 Performance Verification for Digital Generator and Digital Analyzer
64 Keysight U8903B Service Guide
– Digital output sampling rate: 48 kHz
– Digital output function: SINE waveform
– SINE waveform frequency: 1 kHz
– SINE waveform amplitude: 1 FFS
– Digital output state: ON
4 Verification steps for word length at 8 bits:
a Set the DSI audio resolution to 8 bits.
b Set the DSI audio word length to 8 bits.
c Set the DSI Master Clock multiplier to 128.
d Set the digital output sampling rate to 16 kHz.
e Set the DSI audio resolution based on the Word Length as given in “Test record for
DSI master clock output frequency accuracy verification” on page 128.
f Set the DSI audio word length to 32 bits.
g Set the Digital Output Sampling Rate based on the sampling rate (Hz) as given in
“Test record for DSI master clock output frequency accuracy verification” on
page 128.
h Set the DSI Master Clock Multiplier based on the Master Clock Multiplier as given in
“Test record for DSI master clock output frequency accuracy verification” on
page 128.
i Measure the Master Clock Rate twice and average the two measurements.
j Record the average Master Clock Rate, F
measure
in the “Test record for DSI master
clock output frequency accuracy verification” on page 128.
k Compute the master clock error using the following equation. MCLK is the master
clock rate calculated where MCLK = MUL × Sampling Rate (Hz).
Δ
F
error
ppm = [(F
measure
– MCLK) / MCLK]
×
1000000
5 Verification steps for word length at 12 bits:
a Set the DSI audio resolution to 12 bits.
b Set the DSI audio word length to 12 bits.
c Set the DSI Master Clock multiplier to 96.
d Set the digital output sampling rate to 16 kHz.
e Set the DSI audio resolution based on the Word Length as given in “Test record for
DSI master clock output frequency accuracy verification” on page 128.
f Set the DSI audio word length to 32 bits.