PLZ-5W User’s Manual 35
Basic Functions
Setting the Slew Rate
You can set the speed of change when the current is changed.
The slew rate functions in the following cases.
When the setting is changed to change the current value (including the switching function).
When the current value is changed using external control in constant current (CC) mode.
When the current value is changed while the load is on
The slew rate is set according to the current range as an amount of current change per unit of time. The
value is common to rising and falling slopes.
You can set this regardless of whether the load is on or off. This function operates in CC and ARB modes.
1
Press SOURCE, Level, and then Slew Rate.
2
Use the numeric keypad or the rotary knob to enter the slew rate value.
This completes the setting.
The specified slew rate may not be achieved depending on the load cable inductance. If
this happens, reduce the load cable inductance (p.143).
If the load current is small, the specified slew rate may not be achieved (p.146).
If the current changes when the load is turned on, the slew rate may be slower than the
setting.