EasyManua.ls Logo

Kikusui PLZ205W - Synchronizing the Resuming of Sequences; Aborting Synchronized Operation

Kikusui PLZ205W
178 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
72 User’s Manual PLZ-5W
Advanced Functions Synchronized Operation
Synchronizing the resuming of sequences
A sequence will stop at a step with Wait (pre) set to MSync (p.80).
The resuming of a paused sequence (trigger wait) can be synchronized on multiple synchronized PLZ-
5Ws. To synchronize, you first need to create a sequence (p.74).
1
Press Sequence, Transient, and then Initiate on all PLZ-5Ws to be synchronized.
The sequence is executed.
2
Wait for the sequence on all synchronized PLZ-5W to be paused.
The sequence pauses before executing a step with Wait (pre) set to MSync, and “MSync” appears
in the sub-function area.
3
Press MSync on any of the synchronized PLZ-5Ws.
The sequence resumes simultaneously on all synchronized PLZ-5Ws.
Aborting synchronized operation
The synchronized operation setting is cleared when you remove the LAN cables.