1. Prerequired
Page 1
1.a - Working principle
Using keypad, you can activate (or deactivate) a channel, change the measuring range, set the set points and time-
delay...
Principle: the configuration options are accessed via folders and sub-folders (similar to Windows ). Access is
made via a numerical code (full details in this manual).
®
R
AL1 AL2
CODE:
0000
Visual alarm
LED
16-digit 2-line
display
Keypad
Meaning of the keys
To increment a value or a level
To decrement a value or a level
To validate an input
To cancel an input or to return to the previous step
The Class 200 can output either a voltage or a current signal.
Voltage or Current ?
With the on-off switch located on the left
top of the transmitter (when open), you can
choose analogue output 0-10V (voltage)
or 4-20 mA (current)
Down
0-10 V
Up
4-20 mA
1.b - Output signal selection
Class 200 transmitter configuration via keypad
It's extremely unwise to remove the protection tip of our hygrometry
probes as the sensitive element is very fragile even to light contacts.
However, if you have to remove the protection tip, take all possible
precautions and avoid any contact with the sensitive element.
To remove the protection tip, unscrew it or unclip it.
1.c - Protection tip of the sensor
Protection tip
to unscrew
!
Protection tip
to unscrew
Sensitive
element
Sensitive
element