Important
• You must have approval before sending the SMPTE test pattern.
• Sending a SMPTE test pattern directly to the “Hospital Archive” can cause the
WORKSTATION to lock.
• The tolerance of the “AIM density” can increase or decrease to match the customer
specifications and the WORKSTATION.
• Each gray patch displayed must be the size of the PROBE CUP.
10 Send the SMPTE test pattern to each WORKSTATION “destination”.
11 Use the PHOTOMETER to measure the luminance of the gray patches of the SMPTE
test pattern from 0% - 100% on the WORKSTATION MONITOR.
a. Place the PHOTOMETER PROBE on a flat surface to prevent light from being
detected.
b. Energize the PHOTOMETER.
c. Press “UNITS” until the PHOTOMETER displays “fL”.
d. Press “ZERO” until the PHOTOMETER displays “0.00 fL”.