100
Tank sump gasket.
Check sump
fasteners.
Tighten nuts or
replace o-ring. Do
not over tigten nut.
Washer is not level. Check with level Adjust leveling feet.
Loose connection at
hose clamps.
Tighten hose clamp.
Loose connection at
fill valve.
Check water
supply/ fill valve
connection.
Tighten or replace
seal.
Obstruction in drain
line.
Remove obstruction.
Washer overfilling.
Check operation
of fill sensor.
Recalibrate of
replace water level
sensor.
Washer leaks
around door seal
Improper installation.
Check the cabinet
opening to see if it is
aligned properly with
the door, adjust feet.
Unit not level.
Adjust feet to level
unit.
Door latch
Check door latch
for proper sealing.
Adjust hinge pins or
door striker.
Ethernet
connectivity
No Network. Confirm network
Connect to valid
network.
DHCP Enabled:
setting switched off
Check DHCP
Enabled setting
Turn DHCP Enabled
ON.