Chapter 3: Getting Started
Product Service 1-800-522-7658
34
Original instructions
Place those RFID tags back on the primary filters.
Take off the RFID tags that are on your secondary filters only.
Scan in your black RFID tags from your secondary filters.
Place those RFID tags back on the secondary filters.
The “0” should now be the number of filters that you have. If not, then rescan them.
If it is not in the database, it will scan it in. If it is in the database, it will say
“Already in database.
When you have the correct number of filters press “Ok three times.
Reinsert your maintenance card press “Ok.
You should now see “Enter Filter Lifetime”
365D
5M
730MAX
When commissioning a new hood or changing an Electro sensor, always use the
Max of 730 days. If you are changing out old filters, take the Filter __ D you wrote
down from Step 2 and subtract it from 730 to find the new primary filter lifetime.
For example,
The display from Step 2 says “Filter 85 D”
730
- 85
= 645
In the above example, the number you would enter would be 645 for the days.
Arrow down or up to get your number and then press “Ok.”
You will now see:
Filter list
Total replacement
Partial replacement
Press “X” to return to the previous menu.
You should see:
Language
Date and Time
Filters
Alarms and Sensors
Users and Management