Chapter 3: Getting Started
Product Service 1-800-522-7658
33
Original instructions
***If any of these parameters says “No” then press “X” to go back and create it
again. If on the second attempt you do not see all components then recheck all
component connections or you must contact Labconco Product Service.***
If all parameters say “Yes” then press “Ok.” The hood will now reboot.
Step 6 Scan in Filters.
After rebooting, it will show:
Access granted
(Maintenance)
No Filters installed
>Hood Status
Administration menu
Maintenance menu
Arrow down to the “Administration menu” and press “Ok” to select it.
You should see:
Language
Date and Time
Filters
Alarms and Sensors
Users and Management
Network Parameters
Erase History
Ventilation
Arrow to “Filters” and press “Ok to select it.
You should now see:
>Filter list
Total replacement
Partial Replacement
Arrow down to “Total Replacement” press “Ok.
Press “Ok” three times until you see “Record new filter tags” 0/12.
***Note it is important that the tags are not mixed up. The primary RFID tags remain
on the primary filters and the secondary tags remain on the secondary filters.***
Now take off the RFID tags that are on your primary filters only.
Remove your Maintenance card.
Scan in your black RFID tags from your primary filters.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Labconco Protector Echo 180 Series and is the answer not in the manual?

Labconco Protector Echo 180 Series Specifications

General IconGeneral
BrandLabconco
ModelProtector Echo 180 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals