Page #37 background image
Chapter 3: Getting Started
Product Service 1-800-522-7658
32
Original instructions
Step 5 Create new configuration.
After it reboots you will now see:
>Language
Date and Time
Maintenance Menu
Select “Language” by pressing “Ok.
Arrow down to “English” then press “Ok.
Now press “X” to return to the previous menu.
Arrow down to “Date and Time” press “Ok.
You should see:
>Date
Time
Format
Arrow down to “Format” then press “Ok” to set it to “MMDDYY.
If the date and time are correct you may “Save and Exit”
If date and/or time are incorrect, arrow up and down to each menu press “Ok” to
select it and correct it.
Then “Save and Exit” after your changes.
You should now see:
>Language
Date and Time
Maintenance Menu
Arrow down to “Maintenance” and select it by pressing “Ok.
You will see “Hood Integration” press “Ok.
“Create factory CFG” press “Ok.” Wait a few seconds and you will see the
following:
“Discovered Hardware”
Fan(s) the # of fans that your hood has, either an M1, M2, M3, M4, or M5
Electro (S): “Yes”
Semicon (S): “Yes”
Sash Sensor: “Yes”
Temperature Sensor: “Yes”

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Labconco Protector Echo 180 Series and is the answer not in the manual?

Labconco Protector Echo 180 Series Specifications

General IconGeneral
BrandLabconco
ModelProtector Echo 180 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals