EasyManua.ls Logo

Labconco Protector Echo 180 Series - Page 36

Labconco Protector Echo 180 Series
193 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 3: Getting Started
Product Service 1-800-522-7658
31
Original instructions
Step 3 Remove front panel.
Use your ladder and remove the upper front panel so you can access the RFID
filter tags later in this procedure.
Step 4 Delete current configuration.
You should now see on the screen:
>Hood Status
Administration menu
Maintenance menu
Arrow down to “Maintenance Menu” then press “Ok” to select it.
You should now see:
>Hood Integration
Semicon S. Alarm Delay
Electro S. Alarm Delay
Sash Sensor Trig.
Software Releases
Audible Alarm Mgmt.
At “Hood Integration” press “Ok.
Then arrow down to “Delete Factory Cfg” and press “Ok.
You should now see “0000000000. Press “Ok until the cursor is on the fourth zero
from the right.
“0000000000”
Now arrow up once to make a “1” press “Ok.
Arrow up twice to make a “2” press “Ok.
Arrow up three times to make a “3” press “Ok.
Arrow up four times to make a “4.
You will now have “0000001234.
***Note if you make a mistake you can press “X” to move back to the left and
change the entry.***
If it is correct then press “OK.
You should now see “Hood will reboot please be patient.

Table of Contents

Related product manuals