Chapter 3: Getting Started
Product Service 1-800-522-7658
Press “X” again to return to the previous menu.
Step 8 – Scan user cards.
Now you should see:
Language
Date and Time
Filters
Alarms and Sensors
Users and Management
Network Parameters
Erase History
Ventilation
Arrow down to “Users and Management” press “Ok” to select it.
You should see:
>Set
List
Delete
Select “Set” by pressing “Ok” to select it.
Scan in the “user card” that was used at that hood.
Do this by removing the “Maintenance card” to scan in the “user card.”
You should hear a beep and the screen will say “user added.” Now press “Ok.”
Then “X” to return to the previous menu. Remove user card and reinsert
Maintenance card.
Step 9 – Enter network parameters.
Skip this step if you do not have gGuard.
You should now see:
Language
Date and Time
Filters
Alarms and Sensors
Users and Management
Network Parameters
Erase History
Ventilation
Arrow down to “Network parameters” and press “Ok” to select it.
You will now see: >Ethernet
Bluetooth
Press “Ok” to select Ethernet.