Chapter 3: Getting Started
Product Service 1-800-522-7658
DHCP press “Ok” to say “No” for static IP assignment if dynamic IP used.
DHCP should be “yes.”
Refer to your notes from step 2 or use your lab diagram as a reference.
For Static IP enter what you recorded in step 2.
IP:
MSK:
GW:
When you enter the last number of the “GW:” check that you entered the correct
numbers, if it is not correct then you must correct it before proceeding.
If correct, arrow down to “Save and Exit” and press “Ok.”
You will now see Ethernet
Bluetooth
Press “X” to return to the previous menu:
Language
Date and Time
Filters
Alarms and Sensors
Users and Management
Network Parameters
Erase History
Ventilation
***Do not touch the last two items in the menu.***
Press “X” again to return to the previous menu.
Step 10 – Set fan speeds (NOTE: Fans must be off. Turn off now!)
You should now see:
>Hood Status
Administration menu
Maintenance menu
Arrow down to “Maintenance menu” press “Ok” to select it.
You should now see:
>Hood integration
Fan set point
Semicon S. Alarm Delay
Electro S. Alarm Delay
Sash sensor trigger
Software releases